Low-cost, syringe based ion-selective electrodes for the evaluation of potassium in food products and pharmaceuticals

Elzbieta Jarosinska,Emilia Witkowska Nery,Martyna Durka,Marta Podrażka,Julia Wojnowska
DOI: https://doi.org/10.26434/chemrxiv-2024-sc4fj
2024-03-21
Abstract:Over the years, ion-selective electrodes (ISEs) have become routine means of analysis for industrial, environmental, and clinical applications, allowing for rapid, non-invasive, and easy disease monitoring and diagnostics. Their enormous advantages include wide dynamic range, ease of operation and low detector cost. Sets of ion-selective electrodes selective towards various ions can easily be assembled into an electronic tongue system. The additional information gained from other electrodes can help overcome the ISE’s inherent limitations, such as the need for recalibration in more complex samples of variable ionic strength. In this work we developed low-cost ion-selective electrodes made from disposable syringes equipped with plasticized PVC membranes. To characterize the system we have chosen well-studied ionophore valinomycin, and the results in model solutions were comparable to standard ISE electrodes fabricated using commercial bodies. However, quantification of potassium in real samples showed, as expected, that measurement in complex, more concentrated solutions of higher ionic strength, such as beetroot soup, tomato-based food products, and dried fruits, are reproducible but subject to relative error up to 76 %. To overcome this limitation, we have constructed an electronic tongue based on the low-cost syringe electrodes. Different compositions of the array were tested with a series of multivariate algorithms. We have shown that adding a single type of electrode coupled with an ensemble random forest model allows us to quantify potassium in a wide range of food products, and pharmaceutical supplements without recalibration. Compared to a quantification of potassium using a standard calibration curve, the root mean square error of prediction was almost six times lower. We have also shown how the array can be designed to achieve comparable analytical performance using less selec-tive ionophores, such as dibenzo-18-crown-6 instead of valinomycin.
Chemistry
What problem does this paper attempt to address?
The problem that this paper attempts to solve is the challenge of accurately measuring potassium ion concentration in food products and pharmaceuticals. Although traditional ion - selective electrodes (ISEs) are widely used in industrial, environmental and clinical applications, they have limitations when dealing with complex samples (such as solutions with high ionic strength), for example, they need to be recalibrated to adapt to samples with different ionic strengths. These problems limit the application of ISEs in real samples, especially in the precise measurement of potassium ions in food and pharmaceuticals. To overcome these limitations, the author developed a low - cost syringe - based ion - selective electrode and applied it to the selective detection of potassium ions. By using plasticized PVC membranes and common syringes as the electrode body, this new - type ISE is not only low - cost but also easy to prepare. However, when making measurements in complex real - life samples, such as beet soup, tomato products and dried fruits, these electrodes show a relative error of up to 76%. To solve this problem, the author constructed an electronic tongue system based on low - cost syringe electrodes. By testing sensor arrays with different compositions and combining multiple multivariate algorithms (such as partial least squares, multiple linear regression, principal component regression and random forest algorithm), the author found that adding a single type of electrode in combination with an integrated random forest model can achieve the goal of quantifying potassium ions in various food products and pharmaceutical supplements without recalibration. Compared with using the standard calibration curve for potassium ion quantification, the predicted root - mean - square error is reduced by nearly six times. In addition, the author also demonstrated how to design a sensor array to use lower - cost and less - selective ionophores (such as dibenzo - 18 - crown - 6 instead of valinomycin) to achieve the quantification of potassium ions in a wide range of real - life samples. This not only reduces the cost of the electrodes but also improves the accuracy and reliability of the measurement.