Content Determination of Rifabutin by HPLC

Xian Li
2011-01-01
Abstract:Objective To establish HPLC mathod for the content determination of rifabution.Methods Using p-nitrophenol as the internal standard,Lichrospher100(RP-8 5.0μm×250mm) column was used and the mobile phase consisted of acetonitrile-methanol-0.04mol/L KH2PO4(54:10:36,v/v)with the flow rate of 1.0ml/min.The column temperature was set at 40℃,the injection volume was 5.0μl and the detection wavelength was 254nm.Results The linear range of rafibution was 3.75~150.00μgomL-1(r=0.99987) with an average recovery of 100.24%(RSD=0.89%,n=6).Conclusion The method is simple,rapid,accurate and sensitive for the content determination of rifabutin in the quality control and the clincal monitoring.