Determination of serum nevirapine by RP-HPLC

Fan XU,Yu-wang LIU,Xia ZHAO,Gui-li XU,Yi-min CUI
DOI: https://doi.org/10.3321/j.issn:1003-3734.2005.09.026
2005-01-01
Abstract:Objective:To establish a RP-HPLC method to assay serum nevirapine.Methods:The HPLC conditions were as follows: ZORBAX Eclipse XDB-C_8 column (4.6 mm×150 mm, 5 μm), the mobile phase consisted of 15 mmol·L~ -1 (NH_4)_2HPO_4(pH 7.5 adjusted by phosphate acid)-acetonitrile (80∶20) with the flow rate of 1.5 mL·min~ -1 and the detection wavelength at 295 nm. Levofloxacin was used as an internal standard.Results:The linear curve was in the range of 0.05~8 μg·mL~ -1(r=0.999 3).Conclusion:The RP-HPLC method is simple, accurate and reliable.
What problem does this paper attempt to address?