Method Validation of Plasma Concentration Determination of Estazolam by HPLC

Mingguang Li
2011-01-01
Abstract:OBJECTIVE:To establish the method for plasma concentration determination of estazolam,and to confirm optimal determination conditions.METHODS:HPLC method was adopted.The determination was performed on Elite Hypersil ODS2 C18 column with methanol-acetonitrile-water(28:28:54) as mobile phase at flow rate of 1.0 mL·min-1.The column temperature was set at 35 ℃ and detection wavelength was 230 nm.Diazepam was used as the internal standard.The concentrations of estazolam in plasma,alkalinized plasma and serum in patients receiving estazolam were determined.Selected samples and extraction method were verified.RESULTS:Plasma concentration of estazolam was significantly higher than those in alkalinized plasma and serum.In bilateral paired t test,the concentration of estazolam in plasma was significantly different from that in alkalinized plasma and serum(P=0.011 30 and P=0.018 17).The linear range of estazolam was 0.049 4~1.289 6 μg·mL-1(r=0.991 8).The minimum quantitation limit was 0.049 4 μg·mL-1.The average recovery rate was 99.95%~100.79%.The RSD of intra-day and inter-day were less than 10%.CONCLUSION:It is feasible to collect plasma sample of patients for plasma concentration monitoring of estazolam and quantitation analysis of drug poisoning.The method is simple,accurate.