Study on the Performance of All-Solid-state Copper Ion Selective Electrode Based on PEDOT/PSS

Wenpin Zhang,Shuangquan Wang,Dugang Kang,Zhi Xiong,Yong Huang,Huasheng Xiao,Xi Jia,Wei Zhao,Xiaoyang Wu,Yi Xu
DOI: https://doi.org/10.1117/12.2679075
2023-01-01
Abstract:After preparing an all-solid-state copper ion selective electrode (Cu2+-ISE) based on PEDOT/PSS conversion layer via electrodeposition, this paper characterizes the morphology and cyclic voltammetry performance of the PEDOT/PSS film, optimizes the deposition time of the PEDOT/PSS film, and tests the reproducibility, selectivity and acid-base resistance of the all-solid-state copper ion selective electrode. The experimental results show that the prepared all-solid-state copper ion selective electrode has a linear response in the range of 1x10(-6) mol center dot L-1 similar to 1x10(-2) mol center dot L-1, with a response time less than 30 s, and the response slope of 27.8 mV/ decade. The all-solid-state copper ion selective electrode mentioned above can meet the detection standard of national II surface water.
What problem does this paper attempt to address?