Stable Ion-Selective Electrode Based on Screen Printed Electrode for the Pb 2+ Detection Using Hierarchical Polypyrrole-Cupct Nanowires As the Solid Contact

Xianghua Zeng,Yanhua Liu,Waterhouse I.N. Waterhouse,Xiaohui Jiang,zhiming zhang,Liangmin Yu
DOI: https://doi.org/10.2139/ssrn.3971602
2021-01-01
SSRN Electronic Journal
Abstract:In this work, the stable determination of Pb2+ ion was realized by constructing solid contact ion-selective electrodes (SC-ISEs) with a disposable screen-printed electrode (SPE) as the base electrode. The hierarchical polypyrrole nanowires doped with copper phthalocyanine-3,4',4',4'-tetrasulfonic acid (PPy-CuPcT) were prepared as the solid contact layer. The proposed PPy-CuPcT exhibited high double layer capacitance. The resulting SPE/PPy-CuPcT/Pb2+ - ion selective membrane (ISM) showed a Nernstian response in range of 10-6-10-2M (Pb2+ concentration) with a slope of 29.7 ± 0.6 mV/decade. A detection limit of ~3.98 × 10-7 M can be achieved. The potential drift of SPE/PPy-CuPcT/Pb2+ -ISM was only 3.5 μV/s when the reverse current is applied. Furthermore, the fast response (5 - 8 s), insensitive to O2, N2 and light, and high reproducibility and repeatability of the potential values make it very promising for routine analysis and application.
What problem does this paper attempt to address?