FLUORIDE-SPECIFIC ION ELECTRODE METHOD FOR THE DETERMINATION OF FTORAFUR (FT-207) PLASMA LEVEL

Ming Lu,Liu M,Li D
1981-01-01
Abstract:A quick method, by using of a fluoride-specific ion electrode and a pH meter (±1mV), for the determination of the plasma level of Ftorafur(Ft-207)was hereby reported. The sensitivity of the method was 0.038μg F-/ml, which was equivalent to 0.4μg Ft-207/ml. The range of plasma concentration of Ft-207 determined varied from 4~100μg/ml. After intravenous administration of 20 mg Ft-207/kg to rabbit, the plasma concentrations were measured at different intervals. A two-compartment open model for the deposition of free Ft-207 in plasma was proposed. Its pharmaco-kinetio parameters were:K21 = 0.0400 min-1,K12 = 0.0158min-1, Ke = 0.00157min-1. The half-life of plasma distribution was 12.33 min(t1/2(a))and the half-life of plasma elimination was 10.31 hr(t1/2(/3)).It was obvious that the absorption of Ft-207 was fast while its rate of elimination was so slow that within 2 hrs after intravenous injection, a rather high plasma level was maintained.
What problem does this paper attempt to address?