Rapid Determination of 5-Fluorouracil in Plasma Using Capillary Electrophoresis.

HJ Lu,YL Guo,H Zhang,QY Ou
DOI: https://doi.org/10.1016/s1570-0232(03)00033-3
IF: 3.318
2003-01-01
Journal of Chromatography B
Abstract:A rapid, simple and sensitive capillary electrophoresis (CE) method used for the determination of 5-fluorouracil in rabbit plasma is described in the present paper. In this method, samples were simply pretreated by a solvent extraction procedure prior to injection. With a running buffer composed of 30 mM Tris-H(3)PO(4) (pH 7.0) and 5% isopropanol, 5-fluorouracil was easily separated from the external standard alpha-phenethylol as well as other substances existed in the plasma. A linearity of 5-fluorouracil was determined in the range from 0.17 to 42.50 microg/ml with a correlation coefficient of 0.999. A limit of quantitation (LOQ) corresponding to signal-to-noise ratio of 10 was obtained (LOQ=0.08 microg/ml). The method was successfully used for determining the 5-fluorouracil in real plasma samples from rabbits.
What problem does this paper attempt to address?