Accuracy of the Estimations of Respiratory Mechanics Using an Expiratory Time Constant in Passive and Active Breathing Conditions: a Bench Study

Yuqing Chen,Yueyang Yuan,Hai Zhang,Feng Li
DOI: https://doi.org/10.1186/s40001-023-01146-y
IF: 4.981
2023-01-01
European Journal of Medical Research
Abstract:Respiratory mechanics monitoring provides useful information for guiding mechanical ventilation, but many measuring methods are inappropriate for awake patients. This study aimed to evaluate the accuracy of dynamic mechanics estimation using expiratory time constant (RCexp) calculation during noninvasive pressure support ventilation (PSV) with air leak in different lung models. A Respironics V60 ventilator was connected to an active breathing simulator for modeling five profiles: normal adult, restrictive, mildly and severely obstructive, and mixed obstructive/restrictive. Inspiratory pressure support was adjusted to maintain tidal volumes (VT), achieving 5.0, 7.0, and 10.0 ml/kg body weight. PEEP was set at 5 cmH2O, and the back-up rate was 10 bpm. Measurements were conducted at system leaks of 25–28 L/min. RCexp was estimated from the ratio at 75