Continuous Estimation of Airway Resistance in Non-Invasive Ventilation

Yuqing Chen,Yueyang Yuan,Hai Zhang,Feng Li,Xiaohui Wang
DOI: https://doi.org/10.1016/j.resp.2021.103738
IF: 2.821
2021-01-01
Respiratory Physiology & Neurobiology
Abstract:Background: This study aimed to evaluate the accuracy of expiratory time constant (RCexp) to continuously calculate the airway resistance (R-aw). Material and methods: A Respironics V60 ventilator was connected to a lung simulator for modeling different profiles of respiratory mechanics. Results: During assisted ventilation, the respiratory system compliance (C-rs) calculation was always over-estimated in most lung models. The R-aw estimation using the expiratory resistance (R-exp) method was close to the calculated value with the occlusion method during volume-controlled ventilation (VCV). In expiratory flow limitation (EFL) lung models, similar results were obtained in the estimation of inspiratory resistance (R-insp), but different variations were observed in the calculation of the R-exp. The results estimated with RCexp and with dynamic signal analysis had significant variation and accuracy (p < 0.001). Conclusion: The RCexp method is a robust approach to provide real-time assessments of R-insp and R-exp in spontaneously breathing patients during noninvasive ventilation. An underestimation of R-exp was observed in EFL lung models.
What problem does this paper attempt to address?