Modeling and Analysis of Reverse-Fixed Piezoelectric Tube Fiber Scanner for Two-Photon Endomicroscopy

Conghao Wang,xiulei zhang,huilan liu,jianrui ma,jingquan tian,qiang fu,runlong wu,aimin wang,lishuang feng
DOI: https://doi.org/10.1117/12.2652027
2023-01-01
Abstract:Miniature two-photon microscopy combines the two-photon excitation principle and is more lightweight and flexible, allowing it to be applied to freely behaving animals. The piezoelectric tube (PZT) fiber scanner is the key actuated component in miniature fiber-scanning two-photon endomicroscopy (TPEM). In this paper, we use multi-physics field finite element simulation to model and analyze a reverse-fixed PZT fiber scanner for TPEM. The simulation results show that the first two resonant frequencies of the PZT fiber scanner are 163.6 Hz and 757.9 Hz, respectively. At the first two resonant points, the PZT fiber scanner scan range are 0.078 mm and 0.68 mm, respectively. Theoretical guidance for frequency selection of the reverse-fixed PZT fiber scanner is provided by these simulation results.
What problem does this paper attempt to address?