Evaluating the Impact of Different Positron Emitters on the Performance of a Clinical PET/MR System.

Xiangxi Meng,Hui Liu,Hui Li,Shujing Wang,Hongwei Sun,Feng Wang,Jin Ding,Liuchun He,Xin Chen,Lujia Jin,Yun Dong,Hua Zhu,Zhi Yang
DOI: https://doi.org/10.1002/mp.15513
IF: 4.506
2022-01-01
Medical Physics
Abstract:Purpose The positron range and prompt gamma emission are distinctive with different positron emitters. The performance assessment of an integrated PET/MR scanner with these positron emitters is required for related applications, as the magnetic field interferes with the positron propagation. Such an assessment is to be performed on the United Imaging uPMR 790-integrated PET/MR system. Methods The performance measurement methods were modified based on NEMA NU 2-2012, involving F-18, Cu-64, Ga-68, Zr-89, and I-124 as positron emitters. The NEMA IEC phantom was used for evaluations of image qualities. An agarose cap was wrapped around the point source for tissue-simulating spatial resolution measurement. The count rate performance was assessed with selected positron emitters. Images of a 3D-printed Derenzo phantom and representative patients were also acquired. Results The image quality measurement showed that all five positron emitters were suitable for the PET/MR system studied. However, due to the magnetic field, the image of the point source showed an elongated comet-tail feature, which could be eliminated by a tissue-simulating cap. This effect is more obvious in I-124 and Ga-68, due to their long positron ranges. The imaging ability with various positron emitters was further validated with the count rate assessment, the Derenzo phantom, and the clinical images. Conclusions Different positron emitters could be effectively imaged by the PET/MR system tested. The resolution measurement strategy proposed could be applied to measure PET spatial resolution in the magnetic field.
What problem does this paper attempt to address?