Measurement and Evaluation of Quantitative Performance of PET/CT Images Before a Multicenter Clinical Trial

Yanjia Zhu,Caizheng Geng,Jia Huang,Juzhen Liu,Ning Wu,Jun Xin,Hao Xu,Lijuan Yu,Jianhua Geng
DOI: https://doi.org/10.1038/s41598-018-27143-4
IF: 4.6
2018-01-01
Scientific Reports
Abstract:To ensure the reliability of the planned multi-center clinical trial, we assessed the consistence and comparability of the quantitative parameters of the eight PET/CT units that will be used in this trial. PET/CT images were scanned using a PET NEMA image quality phantom (Biodex) on the eight units of Discovery PET/CT 690 from GE Healthcare. The scanning parameters were the same with the ones to be used in the planned trial. The 18F-NaF concentration in the background was 5.3 kBq/ml, while the ones in the spheres of diameter 37 mm, 22 mm, 17 mm and 10 mm were 8:1 as to that of the background and the ones in the spheres of diameter 28 mm and 13 mm were 0 kBq/ml. The consistency of hot sphere recovery coefficient (HRC), cold sphere recovery coefficient (CRC), hot sphere contrast (QH) and cold sphere contrast (Qc) among these 8 PET/CTs was analyzed. The variation of the main quantitative parameters of the eight PET/CT systems was within 10%, which is acceptable for the clinical trial.