Dose Measurement Experiments for Single and Composite Targets in 6 Mev Linear Accelerators

Zhihui Wang,Jiayang Liu,Jiaru Shi,Hao Zha
DOI: https://doi.org/10.18429/jacow-ipac2019-thprb056
2019-01-01
Abstract:The target in electron linear accelerator plays an important role in the production of photon. Targets of different materials and thicknesses have different X-ray yields. In this study, experiments were carried out to measure the dose rates of single targets and composite targets of different thicknesses for 6 MeV linear accelerators utilizing ionization chamber. The electron energy spectrum at the outlet of accelerating tube was detected with magnetic analyser. The experimental results show consistent rules with Monte Carlo simulations. Composite material target of 1.2 mm tungsten and 2 mm copper can deliver 1242 rad/min/100 μA dose rate at 1 meter in front of the target. Dose rates of tungstenrhenium alloy(74%W-26%rhenium) targets were examined too.
What problem does this paper attempt to address?