Evaluation of Nuclear Activation of Tungsten Plates for Future Modeling of a Medical Accelerator

Bin Han,Bryan Bednarz,Yaron Danon,Robert Block,X. George Xu
DOI: https://doi.org/10.13182/nt09-a9246
IF: 1.667
2009-01-01
Nuclear Technology
Abstract:High-energy photons from medical accelerators are used to treat tumors in cancer patients. One consequence is the production of neutrons from photonuclear interactions in the high-Z accelerator components. The release and capture of neutrons produce radioactive nuclei that can irradiate patients and medical personnel. The goal of this study is to develop a method for quantifying the activation of accelerator components using MCNPX. To benchmark this method, we took activation measurements from the irradiation of a series of zinc plates using a 55-MeV electron beam and compared them with MCNPX calculations. The measured cumulative photon-induced activity from Zn-68(gamma, p)Cu-67 interactions in all of the plates was 10.8 MBq, which is in 5.4% agreement with the calculated value of 10.2 +/- 1.1 MBq. Based on these results, a series of simulations were performed in order to optimize the photon- and neutron-induced activity in tungsten for subsequent experiments. The radioactivity from activated short-lived isotopes and subsequent buildup can be significant from repeated accelerator operations during a day. The approach described in this paper is useful in quantifying the origin and the amount of nuclear activation and the buildup of radioactivity.
What problem does this paper attempt to address?