Evaluating activation of the shielding walls of a treatment room using the Monte Carlo method

D.-Y. Lee,J.-H. Kim
DOI: https://doi.org/10.1088/1748-0221/13/05/P05031
2019-10-03
Journal of Instrumentation
Abstract:This study investigates the radiation activation process in a medical linear accelerator, which creates a photon beam with the energy acquired from accelerated electrons. The concrete shielding walls used in conjunction with a medical linear accelerator occupy the largest portion of facility decommissioning costs. Therefore, to evaluate the activation of the shielding wall, this study simulated the operation of a linear accelerator with high-energy photon beams (10, 15, and 20 MV). The results of the simulations showed that the high-energy photon beams produced a large number of neutrons in the areas around the linear accelerator head. Several radionuclides were identified, and their half-lives and radioactivity levels were calculated. Half-lives ranged from 2.62 hours to 3.68E+06 years, and the radioactivity levels of most of the radionuclides were found to satisfy their respective clearance requirements. These results indicate that photon beams of 15...
instruments & instrumentation
What problem does this paper attempt to address?