Abstract:To obtain the accurate magnetic properties of the rare earth permanent magnet with high coercivity, a new system for calibrating a pulsed field magnetometer (PFM) is developed. A set of calibration methods is proposed to calibrate the field and magnetization of the PFM. The field is calibrated by using a Rogowski coil and a calibrated pick-up coil. A measurement system is designed and used to measure the magnetization. The signal of the measurement system is calibrated by using the pure nickel sample and the pure copper sample. The samples of iron, nickel, ferrite and NdFeB are measured, and the experimental results show the PFM calibrated by this system has good measurement performance.