A Calibration Method for Coil Constants Using an Atomic Spin Self-Sustaining Vector Magnetometer

Q. Zhao,B. L. Fan,S. G. Wang,L. J. Wang
DOI: https://doi.org/10.1016/j.jmmm.2020.166977
IF: 3.097
2020-01-01
Journal of Magnetism and Magnetic Materials
Abstract:The self-sustaining atomic magnetometer has the advantage of high sensitivity and long spin coherence lifetime, based on which we demonstrate a novel method to calibrate the magnetic coil constants precisely. Via nondestructive phase measurement and coherent optical pumping, the spin polarization of rubidium atoms is regenerated coherently and the Larmor precession signal is oscillating continuously. In this stable state, the calibrating capability is achieved by applying current to coils and scan the magnetic components along x- and y-direction. The magnetic field magnitude is obtained from precession frequency and the coil constants can be derived from the fitting equation directly. The constants of coils in the experiment are 246.010 +/- 0.034 nT/mA and 197.452 +/- 0.025 nT/mA in the x- and y-directions, respectively.
What problem does this paper attempt to address?