Analysis on the Accuracy of CT-guided Radioactive I-125 Seed Implantation with 3D Printing Template Assistance in the Treatment of Thoracic Malignant Tumors.

Zhe Ji,Haitao Sun,Yuliang Jiang,Yi Chen,Fuxin Guo,Jinghong Fan,Junjie Wang
DOI: https://doi.org/10.1093/jrr/rrab068
2021-01-01
Journal of Radiation Research
Abstract:This article analyzes the accuracy of needle track and dose of a 3-dimensional printing template (3DPT) in the treatment of thoracic tumor with radioactive I-125 seed implantation (RISI). A total of 28 patients were included. The technical process included: (i) preoperative CT positioning, (ii) preoperative planning design, (iii) 3DPT design and printing, (iv) 3DPT alignment, (v) puncture and seed implantation. The errors of needle position and dosimetric parameters were analyzed. A total of 318 needles were used. The mean errors in needle depth, needle insertion point, needle tip and needle angle were 0.52 ± 0.48 cm, 3.4 ± 1.7 mm, 4.4 ± 2.9 mm and 2.8 ± 1.7°, respectively. The differences between actual needle insertion angle and needle depth and those designed in the preoperative were statistically significant (p < 0.05). The mean values of all the errors of the chest wall cases were smaller than those of the lungs, and the differences were statistically significant (p < 0.05). There was no significant difference between the D90 calculated in the postoperative plan and those designed in the preoperative and intraoperative plans (p > 0.05). Some dosimetric parameters of preoperative plans such as V100, V200, CI and HI were not consistent with that of preoperative plans, and the difference was statistically significant (p < 0.05). However, there were no statistical difference in the dosimetric parameters between the postoperative plans and intraoperative plans (p > 0.05). We conclude that for thoracic tumors, even under the guidance of 3DPT, there will be errors. The plan should be optimized in real time during the operation.
What problem does this paper attempt to address?