Dosimetry Verification of Radioactive Seed Implantation for Malignant Tumors Assisted by 3D Printing Individual Templates and CT Guidance.

Zhe Ji,Yuliang Jiang,Fuxin Guo,Haitao Sun,Jinghong Fan,Lujing Zhang,Junjie Wang
DOI: https://doi.org/10.1016/j.apradiso.2016.12.009
IF: 1.787
2016-01-01
Applied Radiation and Isotopes
Abstract:OBJECTIVE:We compared the dose distributions of postoperative plans with preoperative plans for 3D printing template-assisted radioactive seed implantations.METHODS:A total of 14 patients with malignant tumors enrolled in the study. The dose parameters included D90, minimum peripheral dose, V100, V150, and V200. The statistical method was the paired t-test.RESULTS:There was no significant difference in P values between the two groups for all parameters except for V100.CONCLUSIONS:The 3D printing guide template can provide good accuracy for radioactive seed implantation.
What problem does this paper attempt to address?