3D-PRINTING NON-COPLANAR TEMPLATE ASSISTANT CT - GUIDED 125I SEEDS IMPLANTATION ON PELVIC RECURRENT CERVICAL CANCER

P. Jiang,F. Guo,Y. Jiang,Z. Ji,J. Wang,H. Sun,J. Fan
DOI: https://doi.org/10.3760/cma.j.issn.0254-5098.2017.07.003
2017-01-01
Abstract:Objective To investigate the accuracy of 3D-printing non-coplanar template (3D-PNCT) assisted 125I seed implantation with CT guidance in the pelvic recurrent cervical between the preplan and post-plan dosimetric parameters.Methods Nine patients with pelvic recurrent cervical cancer received 125I seed implantation under CT guidance assisted with 3D-PNCT.A pre-plan based brachytherapy treatment planning system (B-TPS) assisted with 3D-PNCT for seed needle depth,direction and angle was designed.The dosimetry parameters including homogeneity index (HI),dose of 90% target volume (Dg0),mPD,volume percent of 100%,150% and 200% prescribed dose V100,V150 and V200 and organ at risk between the pre-plan and post-plan were compared.Results Total seeds number was 675 (median 44,25-114) according to pre-plan,and 669 (median 47,25-113) seeds were implanted actually.138 needles need implant according to preplan,and 132 needles was implanted actually.The median angle deviation was 1.99 ° ± 2.94°(0 °-13 °).There was no significant difference of HI,EI and CI between perand post-plan.The differences of D90,MPD,V100,V150 and V200 between pre-and post-plan were not significant.Conclusions The confidence of pre-plan and post-plan for 3D-PNCT assisted 125I seed implantation in the pelvic recurrent cervical cancer could be accurately performed under CT guidance.
What problem does this paper attempt to address?