RP-HPLC Determination of Cilostazol Tablets and Its Related Substances

Jian WANG,Yu LI,HEN C,Ya-jun DU,Shan ZHENG,Bao-shun ZHANG,Lü-jiang YUAN
DOI: https://doi.org/10.13718/j.cnki.xsxb.2015.12.021
2015-01-01
Abstract:To establish an RP‐HPLC method for the determination of cilostazol tablets and its related sub‐stances ,the separation was performed on an Ecosil C18 (250 mm × 4 .6 mm ,5 μm)column ,with the mobile phase consisting of acetonitrile -0 .07 mol/L phosphate acid solution (adjusted to pH 2 .5 with triethyl‐amine) (38∶62) .The flow rate was 1 .0 mL/min ,the detection wavelength at 257 nm ,the column tem‐perature was 30℃ .Cilostazol and its related substances were completely separated in current chromatogra‐phy system with a good linear range of 2 .5 -100 μg/mL (r=0 .999 9);the average recovery (n=6) was 101 .95% ;the limit of the detection was 1 .6 ng .The related substances met the quality requirements well . T he method is simple ,feasible ,accurate and specific .It can be used for quality control of cilostazol tab‐lets .