Determination of the Related Substances of Cetilistat by HPLC

Libo ZHANG,Bei CHEN,Guping TANG
DOI: https://doi.org/10.13748/j.cnki.issn1007-7693.2017.03.021
2017-01-01
Abstract:OBJECTIVE To establish an HPLC method for determining the contents of the related substances of cetilistat.METHODS The condition of detection were as follows:Agilent Phenyl column (250 rnm×4.6 mm,5 tμm),the gradient mobile phase consisted of 5 mmol·L-1 potassium dihydrogen phosphate (adjust with 10% phosphoric acid to pH 4.0) and acetonitrile with a flow rate of 1.0 mL·min-1,the UV detection wavelength was 226 nm.RESULTS The resolution of cetilistat and relate substances was good.The calibration curves were linear in the range of 0.049 5-1.981 0 μtg·mL-1 for impurity A (r=0.999 9);the calibration curves were linear in the range of 0.059 9-0.399 1 μtg·mL-1 for impurity B (r=0.999 9);the calibration curves were linear in the range of 0.059 6-0.397 5 μtg·mL-1 for impurity C (r=0.999 9).The average recoveries of impurity A,impurity B and impurity C were 104,2%,107.0% and 107.8% respectively;the RSDs were 2.09%,1.77% and 2.18%,respectively.CONCLUSION The method is simple,accurate,reliable and can be used for the quality control of the related substances of cetilistat.
What problem does this paper attempt to address?