Content Determination and Related Substances of Cytidine Disodium Triphosphate for Injection by HPLC

张彤,周长明,纪宏,韩南银
DOI: https://doi.org/10.6039/j.issn.1001-0408.2013.05.25
2013-01-01
Abstract:OBJECTIVE:To establish the method for the content determination of Cytidine disodium triphosphate for injection and related substances.METHODS:HPLC method was adopted.The Thermo Hypersil ODS C 18 column was used with mobile phase composed of reversed phase ion-pair reagent and phosphate buffer at the flow rate of 1.0 ml/min.The column temperature was set at 30 ℃,and the detection wavelength was set at 280 nm.The computation was performed with peak area by external standard method,and the contents of related substance were calculated by self-control method.RESULTS:The linear range of cytidine disodium triphosphate was 90-1 800 mg/L(r=0.999 7)with an average recovery of 102.1%(RSD=1.65%,n=9).The limit of quantity was 7.4 ng.CONCLUSIONS:The method is simple,sensitive and reproducible with sound specificity and reliability,and it can be used for quality control of cytidine disodium triphosphate for injection.