HPLC-UV determination of cefetamet sodium in human plasma and urine
zhang li,qin yongping,nan feng,xiang jin,yu qin,liang maozhi,zhang dan
DOI: https://doi.org/10.13461/j.cnki.cja.006040
2013-01-01
Abstract:Objective To establish a HPLC-UV method for the determination of cefetamet sodium in human plasma and urine.Methods The assay was conducted on a HPLC-UV system consisted of Ultimate XB-C 18(150 × 4.6mm,5μm).The mobile phase were acetonitrile-20mmol/L ammonium acetate(90:10,V/V,pH was adjusted to 5.08 with acetate acid) and acetonitrile-10mmol/L ammonium formate(90:10,V/V,pH was adjusted to 3.53 with formate acid),respectively.The mobile phase was pumped at 1mL/min through the colum.The detection wavelengths was set at 265nm.Plasma sample was first deproteinization by acetonitrile,followed by extraction and removal of the acetonitrile with 2.5mL methylene chloride,Ceftizoxime sodium was used as internal standard(IS).Urine sample was diluted by distilled water and quantified by external reference method.Result For plasma,the calibration curve was linear in the range from 0.125 to 320μg/mL,the lowest limit of quantization was 0.125μg/mL.The recovery of plasma pretreatment was 98.6%~105.9%;the intra-day and inter-day RSD were less than 2.0% and 2.1%,respectively.For Urine,the calibration curve was linear in the range from 2 to 2000μg/mL,the lowest limit of quantization was 2μg/mL.The intra-day and inter-day RSD were less than 3.4% and 4.1%,respectively.Conclusion The methods is simple,rapid,sensitive and accurate for the determination of cefetamet sodium in human plasma and urine.