Determination of Cefaclor in Human Plasma by Reversed-Phase High-Performance Liquid Chromatography with UV Detection and Its Application to the Bioequivalence Studies

Bo Jiang,Zourong Ruan,Honggang Lou,Donghang Xu,Hong Yuan
DOI: https://doi.org/10.1080/00032710903137467
2009-01-01
Analytical Letters
Abstract:A selective and sensitive reversed-phase high-performance liquid chromatography method was developed and validated for quantitation of cefaclor in human plasma using cefradine as an internal standard. Calibration curve was linear over range of 0.1-20 mg . L-1. The intra- and inter-run relative standard deviations of the assay were less than 7%. The mean absolute recoveries determined at the concentrations of 0.3, 3.0, 8.0, and 15.0 mg . L-1 were 69.9%, 69.9%, 77.1% and 72.0%, respectively. The analytical method established was proved to be specific, precise, sensitive, and suitable for applying in the pharmacokinetic and bioequivalence studies of cefaclor in human.
What problem does this paper attempt to address?