The Influence of Detector Spectral Response on the Peak Position Determination in the Spectrometer Wavelength Calibration Process

黄星月,杨怀栋,何庆声,田禾,金国藩
DOI: https://doi.org/10.3321/j.issn:1002-1582.2009.03.001
2009-01-01
Optical Technique
Abstract:The spectral response of a photoelectric detector varies with the light wavelength.This could change the spectral line shape and cause an offset when determining the peak position in the spectrometer wavelength calibration process.Considering the spectrometer with an array detector,a Gaussian-shaped calibration spectral line's integral sampling model is set up.The condition on which the peak position changes upon the impact of the detector spectral response is studied.And the limitation for the detector spectral response not to shift the peak position to another pixel is given.The limitation's mathematical expression for a spectrometer with a grating is deduced.To guarantee adequate wavelength calibration accuracy,in spectrometer design,this limitation could be applied to choose a proper detector.In spectrometer calibration,it could be used to choose suitable spectral lines for wavelength calibration.
What problem does this paper attempt to address?