Calibration of Dual-Channel Raman Spectrometer via Optical Frequency Comb

Shengyujie Lv,Xiaoping Lou,Qiaona Gai,Taotao Mu
DOI: https://doi.org/10.3390/s24041217
IF: 3.9
2024-02-15
Sensors
Abstract:The portable Raman spectrometer boasts portability, rapid analysis, and high flexibility. It stands as a crucial and powerful technical tool for analyzing the chemical composition of samples, whether biological or non-biological, across diverse fields. To improve the resolution of grating spectrometers and ensure a wide spectral range, many spectrometer systems have been designed with double-grating structures. However, the impact of external forces, such as installation deviations and inevitable collisions, may cause differences between the actual state of the internal spectrometer components and their theoretical values. Therefore, spectrometers must be calibrated to establish the relationship between the wavelength and the pixel positions. The characteristic peaks of commonly used calibration substances are primarily distributed in the 200–2000 cm−1 range. The distribution of characteristic peaks in other wavenumber ranges is sparse, especially for spectrometers with double-channel spectral structures and wide spectral ranges. This uneven distribution of spectral peaks generates significant errors in the polynomial fitting results used to calibrate spectrometers. Therefore, to satisfy the calibration requirements of a dual-channel portable Raman spectrometer with a wide spectral range, this study designed a calibration method based on an optical frequency comb, which generates dense and uniform comb-like spectral signals at equal intervals. The method was verified experimentally and compared to the traditional calibration method of using a mercury–argon lamp. The results showed that the error bandwidth of the calibration results of the proposed method was significantly smaller than that of the mercury–argon lamp method, thus demonstrating a substantial improvement in the calibration accuracy.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The problem this paper attempts to address is: how to improve the calibration accuracy of dual-channel portable Raman spectrometers over a wide spectral range. Traditional calibration methods (such as using a mercury lamp) have the issue of uneven distribution of characteristic peaks over a wide spectral range, leading to significant errors in polynomial fitting results. To solve this problem, this study proposes a calibration method based on an Optical Frequency Comb (OFC), which generates dense and evenly distributed comb-like spectral signals, thereby significantly improving calibration accuracy. Specifically, the main contributions of this paper include: 1. **Proposing a new calibration method**: Generating evenly spaced dense spectral signals based on an Optical Frequency Comb to address the issue of uneven distribution of characteristic peaks in traditional calibration methods. 2. **Designing a dual-channel calibration process**: Calibrating the high wavenumber and low wavenumber channels separately to ensure calibration accuracy across the entire spectral range. 3. **Experimentally validating the effectiveness of the method**: Demonstrating a significant improvement in accuracy of the OFC-based calibration method compared to the traditional mercury lamp calibration method. These improvements are of great significance for enhancing the reliability and accuracy of Raman spectrometers in practical applications.