Study on the effective dose for patient receiving PET-CT under various fixed tube currents mode

梁子威,耿建华,赵葵,王奕斌,杜召猛,毕高畅,齐雍鹤,吴宁,郑容
DOI: https://doi.org/10.3969/J.ISSN.1672-8270.2018.04.003
2018-01-01
Abstract:Objective: To research the effective dose for patient receiving PET-CT under fixed tube currents mode so as to provide reference bases for determining optimum clinical scan scheme. Methods: The PET-CT of Discovery Elite was adopted as experiment equipment, and the effective dose of CT scan was obtained from simulated experiment for simulation phantom(Model RS-550). Collected conditions of CT included that the current range of fixed tube was 30-250 mA with an interval of 20 mA, and tube voltage was 120 kV, and screw pitches were 0.984 and 1.375, and the rotate speeds were 0.5 s/R and 0.8s/R. The segmented scan mode that simulated clinical PET-CT was implemented for whole body of phantom. Besides, the dose-length products (DLP) under various scan conditions were recorded, and the effective dose was further calculated. Results: The range of effective dose caused by CT under fixed tube current mode was between 1.05 and 18.99 mSv. And with increasing of current and rotate speed, the effective dose increasing linearly, while with increasing of screw pitches, it decreasing linearly. Conclusion: For a certain patient, there is a noteworthy phenomenon is the difference of effective dose among various collecting conditions was larger under fixed tube current scan mode, and the effective dose of Discovery Elite is lower than that of Discovery ST-16. Therefore, according to different clinical requirement, flexible choosing different CT scan condition that aimed at different machine type can effectively reduce effective dose for patient under the precondition of ensuring image quality.
What problem does this paper attempt to address?