A radiation feature study and dose safety management suggestion on Philips Multislice CT iCT 256 under axial scanning mode

Xiaoli LIU,Hailiang ZHENG,Xingdong LI,Shuli FENG,Rengui WANG,Yongli DUAN
DOI: https://doi.org/10.3969/j.issn.1002-3208.2014.03.12
2014-01-01
Abstract:Objective To verify the reported CTDIvol value and explore dose safety managementprotocol of Philips Brilliance iCT 256 under axial scan mode with head protocolMethods A pencil dosimeter,Solidose 400 from Sweden RTI Electronics,is utilized to detect radiation dose index of phantom T6M1 64 (incompliance with IEC standard 61 223-2-6,provided by National Institute of Metrology of China)when scannedby iCT 256 under head axial protocolResults (1)Relative deviation of CTDI1 00 in the center point is about 2%and those in the peripheral 4 points are about 20% (2)Tested CTDIvol value is 5% ~1 6% greater thanreported value(3) Effective absorbing dose during one head axial scanning is about 086 mGy to 363mGyConclusions Accuracy of CTDI1 00 measurement in the center point is better than the peripheral fourpointsTested CTDIvol value might be greater than reported CTDIvol in certain conditions and supervision onaccumulated radiation dose of patient should be recommended.