Effect of Offset Distance between X-Ray Beam and Ionization Chamber Center to CTDI100 Radiation Dosage Measurement of Multi-Slice CT

Hai-liang ZHENG,Shu-li FENG,Ren-gui WANG,Yong-li DUAN,Lei YANG,Yan FU,Jun ZHAO
DOI: https://doi.org/10.3969/j.issn.1674-1633.2012.09.006
2012-01-01
Abstract:Objective To explore the influence law of offset distance(Do) between ionization chamber center scale and X-ray beam plane of multi-slices CT to the CTDI100 measurement result.Methods Head body mold(T6M164) and long pole ionization chamber,which meet IEC standards,are used to measure the CTDI100 value of Philips Brilliance iCT 256 on axial scanning condition of different offset distance.Results when Do is 1cm,relevant deviation(Dr) is 0%~5%,when Do is 3cm,Dr is 5%~15%,and when Do is 5cm,Dr is 25%~75%.Conclusion The measurement result of CTDI100 has greater sensitivity for Do,which means the operator should make sure no deviation exist between X-ray beam and ionization chamber center,besides,more attention should be paid to radiation dose in penumbra area.Scattering beam causes the measurement of CTDI100 at 4 points around more discrete,effect of bed to measurement result shows that symmetry of CTDI100 at 4 points outside the center reduces.