Dosimetric consistency of different Varian kV-cone beam CT systems: a comparative study

Meijiao WANG,Yibao ZHANG,Sha LI,Fei JIA,Hao WU,Yanqiu DING,Wen GUO
DOI: https://doi.org/10.3969/j.issn.1005-202X.2017.06.002
2017-01-01
Abstract:Objective To verify the dosimetric consistency between two Varian CBCT systems of the same type but of different working years using 6 sets of standard scanning parameters.Methods The stability of the tube voltage (kVp) of the two systems was measured using PTW DIAVOLT-MULTI (Germany).In a standard CTDI phantom consisting of the head and body modules (diameters of 16 cm and 32 cm,respectively),the dose at the central and 4 peripheral points were measured with a PTW CT chamber (Germany) to calculate the weighed CT dose index (CTDIw) in both the full-fan and half-fan modes.Results Measurement using the 6 sets of standard scanning parameters showed that the kVp of both the new and old CBCT systems was higher than the nominal values by (2.5±0.7) kVp and (2.9±0.8) kVp,respectively,with deviations ranging from 1.3% to 3.12%.Compared to the new system,the old CBCT system had a higher CTDIw by (8.73±2.95)% on average,with deviations ranging from 4.90% to 11.46%.Conclusion The dosimetric consistency of Varian CBCT is stable but the deviation of kVp increases as the working year of the machine increases.The CTDIw can vary between different Varian CBCT systems even of the same type and with the same scanning parameters.Monitoring of the dosimetric consistency is recommended for routine quality control and maintenance of the CBCT systems.