The clinic application of retained guidewire in left ventricular lead-12 cases

蓝荣芳,吉文庆,狄文成,李晓宏,余洪松,王如兴,顾翔,杜荣增,徐伟
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2014.05.008
2014-01-01
Abstract:Objective To investigate the safety of retained guidewire in left ventricular lead.Methods The retained percutaneous coronary intervention(PTCA) guidewire technique had been used in 12 patients for repetitive intraoperative dislocation.The threshold and the impendence of the left ventricular lead were recorded during follow-up.The patients with abnormal parameters took chest X-ray.Results Lead fracture was demonstrated by X-ray 6 months after implant in 1 patient.The marked increase in pacing threshold was recorded one month after implant in 2 patients with phrenic nerves stimulation as well as guidewire fracture.There was pacing threshold increase in 2 patients without changes of impendence and X-ray for 1 to 3 years.There were no changes in threshold,impendence as well as X-ray in 7 patients for 1.5 to 4.0 years.Conclusion Retained guidewire technique should be forbidden because it can damage the lead.