Lin Ruowei,Li Xiaohong,Ji Wenqing,Xu Wei
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2015.02.009
2015-01-01
Abstract:Objective Observe the influence of retained guidewire on the pacing lead.Methods From Jan.2010 to Dec.2013,Nanjing Drum Tower Hospital,six patients received leads extraction because of lead fracture were enrolled in our study.Results Five patients had retained guidewire.All leads were extracted using mechanical tools,including 2 right ventricular fixed leads and 3 left ventricular pacing leads.Fracture of the coil electrodes as well as deterioration of the insulation coating was found in 4 patients.1 retained guide in left ventricular pacing fracture,but the pacing parameters,including impedance was not elevated.Conclusion The retained guidewire technique should not be used because delayed electrode damage can occur.