Application of a novel pacing guide wire in cardiac resynchronization therapy

YU Hai-bo,LIANG Yan-chun,WANG Zu-lu,SUN Yi,XU Guo-qing,JIN Zhi-qing,LI Shi-bei,FU Liu-jing,HAN Ya-ling
DOI: https://doi.org/10.1136/heartjnl-2013-303992.035
IF: 5.7
2013-01-01
Heart
Abstract:Objective To investigate the availability of a novel Visionwire guide wire in the implantation of left ventricular(LV) lead in cardiac resynchronization therapy(CRT).Methods Five heart failure patients selected for CRT were involved in the study.The pacing threshold,R-wave sensing,phrenic nerve stimulation at local coronary sinus(CS) branch were measured by the Visionwire guide wire and LV lead separately.Results The pacing parameters were analysed through the Visionwire guide wire and LV lead separately at the 16 CS branches in 5 patients.There were no significant difference between the pacing parameters measured by the two methods.Significant correlation was found between LV pacing threshold assessed by the Visionwire guide wire and LV lead(r=0.90,P<0.01).Correlation for R wave sensing was also significant(r=0.67,P<0.01).The pacing sites accompanied with phrenic nerve stimulation while pacing at 10 V/0.42 ms were similar by the two methods.Visionwire guide wire could be used for local potential electrical mapping.The procedure time for pacing test in a single CS branch by Visionwire guide wire was much less than that by LV lead(12.4±7.5 min vs 18.3±12.2 min,P<0.01).Conclusions Visionwire guide wire facilitates transvenous LV lead implantation by prediction of pacing parameter and locate the target CS branch quickly.