The effect of distance from light source on the light intensity decrease of six LED light units
XIAO-dong XU,XIANG-feng MENG,RONG-rong NIE
DOI: https://doi.org/10.15956/j.cnki.chin.j.conserv.dent.2014.11.011
2014-01-01
Abstract:AIM:To investigate the effect of distance from light source on the light intensity decrease of 6 LED light units.METHODS:The light intensity LED light units,Bluephase C8,Bluephase and Elipar S10 equipped with standard light guide,Ascent PX equipped with a condenser lens,Demi and Ti-Lite GT-1500 equipped with a tur-bo light guide,was respectively measured at 0,1,2,3,4,and 5 mm respectively from the tip of the light guide with a radiometer.Measurements were repeated for 10 times at each distance to obtain the mean value and the decrease per-cent of the light intensity.RESULDS:The light intensities at 0 mm from light source were 859 mW/cm2 (Bluephase C8),1 121 mW/cm2(Bluephase),1 133 mW/cm2(Elipar S10),1 300 mW/cm2(Ascent PX),1 497 mW/cm2(De-mi)and 1 192 mW/cm2 (Ti-Lite)respectively,and the light intensity decreased for all LED light units with the in-crease of the distance from light source (P<0.05),except for Ascent PX at 1 -3 mm from light source of.At 5 mm, the light intensity of Bluephase,Elipar S10,and Ascent PX were more than 400 mW/cm2 ,and their according de-crease percent of light intensity were 56.7%,58.7%,and 53.8% respectively,which were significantly lower than that of Bluephase C8 (66%),Demi (78.8%)and Ti-Lite GT-1 500 (>72.2%).CONCLUSION:High power LED unit with an extensive exposure time is necessary to compensate the attenuation of light intensity with the increase of tooth cavity depth.High power LED unit with a standard light guide or a condenser lens is recommended for clinical application.