A Development of 3 - D Image Survey System for Tooth Wear Measurement

潘洁,王嘉德,王勇
DOI: https://doi.org/10.3969/j.issn.1003-7632.2005.01.002
2005-01-01
Abstract:Objective To develop and evaluate a noncontact 3D image survey system to measure tooth wear. Methods The dental cast was moved horizontally by a quantifiable manual control platform. The occlusal surface 3D coordinate of each molar (3000-4000 points under the condition of 0.05 mm step) were obtained through the light beam and a special Charge Coupled Device (CCD). The collection and display of the images, as well as all measurement were performed and completed under the Windows system. Results The system offered 15mm x 15mm x 15mm measurement scope. The system's accuracy was 0.01 mm with a mean deviation of less than 0.01mm. A group of standard samples was measured by the new system, and a universal microscope, respectively. There was no significant difference between the two methods ( P 0.05). Conclusion Compared with traditional methods, the new system demonstrates superior precision and several other advantages as noncontact measurement, including more stability, simplicity, and higher efficiency.