An evaluation on influence factors of radiation dose for the medical X-ray computed tomography
LU He-qing,ZHU Guo-ying,ZHUO Wei-hai,GUO Chang-yi,GAO Lin-feng,LIU Hai-kuan
2008-01-01
Abstract:Objective To evaluate those factors which affect on to the radiation dose of multi-slice CT,such as voltage and current of tube,diameter of phantom,depth of inside phantom,pitch,collimation,field of view,algorithm and tube current modulation technology and so on.Methods A pencil-like ion chamber,head and body dose phantom,and Chinese anthropomorphic phantom were used to measure the characteristic quantum of CT dosimetry when different voltage and current of tube,diameter of phantom,depth of inside phantom,pitch,collimation,field of view,algorithm and tube current modulation technology were adopted.Results When all other parameters kept unchanged,the tube voltage was direct proportional to CTDI(CT dose index),the tube current was positive linear correlation to CTDI,the diameter of phantom was inverse proportional to CTDI,radiation dose of CT was decreasing while measuring depth increased of inside phantom,pitch was inverse proportional to accumulated dose,overexposure from spiral scanning out of topogram range led to the increasing of radiation dose,collimation was inverse proportional to CTDI,CTDI was not affected by field of view and algorithm,and radiation dose of CT was decreased by using tube current modulation technology.Conclusion Radiation dose of CT can be decreased by reducing voltage and current of tube,increasing pitch,limiting the block number of helical scanning during a single CT examination,increasing width of collimation,decreasing scan parameters of slim patients rationally and adopting tube current modulation technology.