A preliminary study of exposure dose reduction in full-field digital mammography

张静,汪登斌,王媛,李惠民,王丽君,江怡菲
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2015.07.011
2015-01-01
Abstract:_Objective:To explore the radiation dose in full-field digital mammography and find out the impact factors in image quality and average glandular dose (AGD)to provide valuable exposure parameters setting for mammography. Methods:The NA18~220 mammography phantom was exposured by GE Senography DS full-field digital mammography with the standard (STD)mode of automatic optimization of parameter (AOP)system (pressure:12daN;thickness:42mm). Three exposure parameters (28kV,45mAs,target/filter:Mo/Rh)were acquired as a reference.Firstly,tube current as a variable parameter was ajusted from 80mAs to 90mAs.Then,tube potential wass increased step by step from 22kV to 36kV with 1kV interval.All images obtained from three groups were evaluated by three radiologists according to the criterion of American college of radiology.The number of calcification,nylon fibers and masse shown on the image and the value of AGD were recorded.Results:Compare to AOP system,the image quality was clinically acceptable in the range of 3kV dec-reace in tube potential and 20mAs decrease in tube current,the radiation dose was optimized effectively.The increase in tube voltage potential were prone to achieve better image quality and rising of AGD,compared with increase in tube current.Con-clusion:To maintain the acceptable image quality,the reduction in radiation dose can be achieved by reasonable decreasing tube current and tube potential.