A Study of Correlation between Quality and Exposure of Chest Digital Radiography

ZHAO Xi,YANG Shuo-hui,YU Jian-li,WANG Guo-nian,ZHAN Song-hua
DOI: https://doi.org/10.3969/j.issn.1674-4659.2009.01.003
2009-01-01
Abstract:Objective:To investigate the correlation between the exposure and quality of chest digital radiography.Methods:We retro-spectively analyzed the chest radiography of 170 patients with chest study in our hospital from February 12,2008 to March 2,2008,mea-sured and recorded the background signal value,the signal value of the central of 9th thoracic vertebrae,liver and lung field,as well as kV,mAs,ms,EI,dGycm2.Results:The average radiation dose was 0.555 ± 0.126dGycm2,the average background signal was 26,895 ± 1,079,the correlation coefficient between of them was 0.761(P<0.05).Liver /background signal was associated with radiation dose(P<0.05),and the correlation between lung field /background signal and the 9th thoracic/background signal with radiation dose were not sig-nificant(P>0.05).The mAs and ms were closely related with the radiation dose.All of the parameters were in the range,which were pro-vided by the equipment manufacturers.Conclusion:It's feasible to evaluate the quality of chest radiography by measuring the signal value of selected points in DR imaging.There is a significant correlation between the contrast of chest radiography and automatic exposure dose.Applying the selection of automatic exposure conditions could meet clinical requirements.To the lean patients,automatically re-ducing the radiation dose has little effect on the image quality.