Radiation Dose Optimization of DR Chest Bedside Photography:Experimental Study

唐辉,殷焱,路青,王嵇,朱国英,庄治国,俞璐,马鋆,杨咏琰,许建荣
DOI: https://doi.org/10.13437/j.cnki.jcr.2011.03.001
2011-01-01
Abstract:Objective To investigate the optimal radiation dose of DR chest bedside photography.Materials and Methods The chest model underwent DR chest bedside photography by two methods,increasing KV with fixed mAs and increasing mAs with fixed KV respectively,using Shimadzu Mobile DaRt beside machine.The radiation dose was measured by a radiometer.Each group of images was scored by two experienced radiologists by double blind method.Enumeration data were analyzed by t test,and categorical data were analyzed by chi squared test.Results When mAs was fixed and KV increased from 70kv to 85kv,the area under ROC curve(Az) corresponding to the image quality of each group was 0.675,0.383,0.733,and 0.250,respectively.When KV was fixed and mAs increased from 0.5mAs to 1.8mAs,the area under ROC curve(Az) corresponding to the image quality of each group was 0.817,0.650,0.683 and 0.725,respectively.Conclusion Either voltage or current will affect the radiation dose.Both of them are related to the radiation dose.Appropriate exposure conditions should be selected in clinical application.