Establishment and performance evaluation of LC-MS/MS for the determination of plasma paclitaxel
Kouqiong WANG,Wenqi SHAO,Yingfei PENG,Fei HUANG,Fangjun CHEN,Jiong WU,Beili WANG,Chunyan ZHANG,Wei GUO,Baishen PAN
DOI: https://doi.org/10.3969/j.issn.1673-8640.2016.05.012
IF: 1.865
2016-01-01
Laboratory Medicine
Abstract:ObjectiveTo establish a liquid chromatography-tandem mass spectrometry(LC-MS/MS)for the quantitative determination of plasma paclitaxel,and to evaluate its performance.Methods Plasma paclitaxel was determined quantitatively by LC-MS/MS. High performance liquid chromatography was used. The flow speed was 0.3 mL/min. Multiple reaction monitoring(MRM) was used for quantitative determination. The ion pairs of paclitaxel and PTX-D5 were 876.4>308.2 and 881.5>313.2,respectively. According to U.S. Food and Drug Administration(FDA) guideline about bioanalytical method validation,the general analytical performance validation was performed for linearity,detection limit,precision and accuracy.ResultsThe linearity was 10-1 000 ng/mL. The detection limit was 10 ng/mL. Within-run and between-run coefficients of variation(CV) were≤4.2% and≤8.9%,respectively. The accuracy was 88.7%-104.0%.Conclusions The performance of established LC-MS/MS meets evaluation standards,and the assay is accurate and precise for the determination of paclitaxel.