Determination of 6-mercaptopurine in human plasma by RP -HPLC

Yan-yan ZHU,Zhi-neng BAI,Li-qin TANG
DOI: https://doi.org/10.3969/j.issn.1009-6469.2016.06.015
2016-01-01
Abstract:Objective To establish a reversed phase high -performance liquid chromatography(RP -HPLC)method to determine 6-mercaptopurine (6-MP)in human plasma,thereby providing the foundation for clinical blood concentration monitoring.Methods U-sing Hypersil ODS2 column (250 nm ×4.6 nm,5 μm),the mobile phase was acetonitrile -0.3% glacial acetic acid (3∶97),the flow rate was 0.8 mL · min -1 ,the detection wavelength was set at 323 nm,and plasma samples were extracted by 20 μL 15% perchloric acid.Results A linearity was observed within the rangefrom 0.02 mg ·L -1 to 2.00 mg ·L -1 ,the lowest detection limit was 0.02 mg ·L -1 ,the average method recovery rate of four concentrations were 85.71%,92.93%,103.53% and 96.02%,respectively,and the intra -day and inter -day precision are <15% and in line with the Chinese Pharmacopoeia biological sample analysis regulations.Con-clusions The experiment successfully established a simple,fast,and low cost 6-MP plasma concentration measuring method with high sensitivity and good stability.Particularly,the method needs less blood.It is also in line with the relevant provisions of the AZA blood drug concentration monitoring and pharmacokinetics study.
What problem does this paper attempt to address?