Comparison of Headspace-Gc with Purge and Trap-Gc/ms Methods in Determination of Acetaldehyde, Acrolein,acrylonitrile and Pyridine

LUO Yi,JIANG Jianbiao,CHANG Qing
DOI: https://doi.org/10.7535/hbgykj.2017yx03010
2017-01-01
Abstract:In order to further improve the efficiency and accuracy of determination of acetaldehyde,acrolein,acrylonitrile,pyridine in water.The detection limit,precision and recovery of acetaldehyde,acrolein,acrylonitrile and pyridine in water by gas chromatography with purge and trap GC/MS methods are compared.The precision of headspace-gas chromatography is between 1.4% ~ 3.8%,and the recovery rate is between 93.2% ~ 101.0%.The precision of purge and trap-GC/MS is between 1.7% ~3.2%,and the recovery rate is between 92.3%~99.8%.The linearity of the two methods is good,and the purge & trap-GC/MS method has a lower detection limit.The result shows that the two methods can accurately determine acetaldehyde,acrolein,acrylonitrile and pyridine in water.The sensitivity of purge and trap-GC/MS method is higher,suitable for the determination of low concentration samples;the headspace-gas chromatography is suitable for the monitoring of high concentration or dirty samples.
What problem does this paper attempt to address?