Determination of Trichloroacetaldehyde in Water by P&T-GC-MS

Jiang Peixia,Liu Yong,Chu Chunying,Peng Changsheng
DOI: https://doi.org/10.3969/j.issn.1673-1212.2012.06.035
2012-01-01
Abstract:P&T(Purge and Trap)-GC/MS method was used to determine trichloroacetaldehyde in water.The influence of purge time and NaOH dosage were studied.The results show that purge time is 4 min and NaOH solution dosage of 0.20 ml to 40 ml of water sample,when standard curve 1.0~200 μg/L,linear correlation coefficient above 0.999,sample amount 5 ml,detection limit 0.21 μg/L.Precision and accuracy results show that when standard solution concentration to blank water increased to 2.0~20 μg/L,the average recovery rate of six parallel tests is 108%~94.4%,and RSD is 3.1%~2.4%.
What problem does this paper attempt to address?