Determination of Trace Acrylonitrile in Industrial Wastewater by Headspace-Gc(fid)

JIANG Junkang,DING Shaohong,QU Jianhua,WANG Chun,CHEN Gang
DOI: https://doi.org/10.3969/j.issn.1006-2440.2011.05.005
2011-01-01
Abstract:Objective: To establish a determination method of trace acrylonitrile in industrial wastewater by GC(FID).Methods: Acrylonitrile in water was segregated by headspace bottle in tepidarium into the gas,detected by gas chromatography with hydrogen flame ion detector(GC-FID),and the optimal parameters of the method were explored(headspace,balance temperature and time,and chromatographic condition).Results: Under the optimal conditions above,the retention time of acrylonitrile was 3.60 minutes,linear range of this method was from 1.26mg/L to 20.15mg/L,regression equation was y(areas of chromatographic peak) =0.314+2.60x(concentration of standard),correlation coefficient r=0.9996,detecting limit was 0.08mg/L,relative standard deviation was between 2.66% and 4.63%,recovery of added standard was between 97.67% and 104.78%,and acrylonitrile did not be interfered with some coexistence substances: methanol,n-propanol,acetonitrile and benzene.Conclusion: The determination method has been proven to be simple and effective,with good accuracy,precision and sensitivity,and can satisfy the monitoring demand of acrylonitrile in water.
What problem does this paper attempt to address?