Hand-Held Multi-Channel Detector Based on Amperometric Biosensors for Biochemical Measurement

LUO Xian-bo,JIANG Li-ying,LI Hua-qing,LIUA Chun-xiu,WUC Yi-rong,CAI Xin-xia
DOI: https://doi.org/10.3969/j.issn.1004-1699.2005.03.009
2005-01-01
Abstract:With immobilized dissimilar enzymes or especial charge-transfer mediator on the surface of the(electrode),the multi-channel detector can be used for measurement of a variety of analyzes.The measurement of glucose,lactate and hemoglobin was demonstrated.The measurement range of glucose is 1~25 mmol/L,the correlation coefficient is 0.989;The measurement range of lactate is 1~15 mmol/L,the correlation coefficient is 0.988.In the hemoglobin concentration range of 0~20 g/dL,the measurement current value is correlated to the concentration of hemoglobin.