Study on β-hydroxybutyrate electrochemical biosensor

Nian-zhang MA,Guang LI,Gui-quan HU
DOI: https://doi.org/10.3785/j.issn.1008-973X.2005.03.003
2005-01-01
Abstract:β-hydroxybutyrate electrochemical biosensor was prepared to determine the concentration of serum β-hydroxybutyrate. With potassium ferricyanide as electron transfer mediator, β-hydroxybutyrate dehydrogenase (HBDH) and cofactor β-nicotinamide adenine dinucleotide (NAD) were immobilized on a screen-printed carbon electrode by direct adsorption on the electrode surface during the fabrication of this biosensor. Amperometry was used for determining the β-hydroxybutyrate concentration, and the concentrations of potassium ferricyarride and NAD, pH and temperature were optimized. The experimental results show that the response currents measured by this biosensor have good linear relationship with β-hydroxybutyrate concentration in the range of 30~550 mg/L, have high precision and good repeatability, and what is most important is that only 2 μL sample is needed.