Properties of poly-β-aminoanthraquinone modified carbon fiber electrode as a basis for hemoglobin biosensors

Huangxian Ju,Hongbin Sun,Hongyuan Chen
DOI: https://doi.org/10.1016/0003-2670(96)00067-0
IF: 12.545
1996-01-01
Biosensors and Bioelectronics
Abstract:An amperometric microsensor for hemoglobin (HB) has been constructed using a carbon fiber electrode modified with poly-(pAAQ). Scanning electron microscopic studies showed that the potential range and time of electrochemical polymerization had strong effects on the morphology of the polymer film at the microelectrode. The stability and the electrochemical properties of the modified electrode were studied using cyclic voltammetry. The pAAQ modified microelectrode showed an excellent electrocatalytic activity for the reduction of HB in a weakly acidic solution. The catalytic current in cyclic voltammetry increased linearly with the HB concentration between 1.0–400 μM. The amperometric response to HB had a linear range of 0.5 μM to 340 μM with a correlation coefficient of 0.990(n = 18). The biosensor was stable for at least one month when it was stored in a phosphate buffer (pH 7.4) below 20 °C. The obtained HB concentration in the whole blood agreed well with the one obtained with the reference instrument.
What problem does this paper attempt to address?