Abstract:A double-chamber microbial fuel cell (DMFC) toxic sensors was constructed to detect Cr (Ⅵ),Cd2+,Cu2+,Zn2+,respectively.After optimizing the reactor performance,DMFC reached the maximum power density when the external resistance was 1 000 Ω,and the external circulation rate was 0.933 4 mL · min-1,It was determined that the best detection time was 60 min and the best cleaning time was 10 min by copper detec ting under optimal conditions.The results indicated that the detection ranges were from 0.4 to 10,0.5 to 10,40 to 160,15 to 80 mg · L-1 for four metals respectively in which the initial concentration was defined as the detection limit.Inhibition rate was used to verify the feasibility of the sensors.Within the detection range,we obtained four linear equations and correlation coefficients were 0.959 7,0.979 5,0.944 1,0.936 6.The concentrations within the detection range were selected to verify the stability of the sensors and the relative errors of detection for four heavy metals were less than 11%,which indicated that the sensor was stable and can run for a long time.