Flow field CFD analysis of axial flow blood pump

XIE Xiong,TAN Jian-ping
DOI: https://doi.org/10.3969/j.issn.1001-3881.2014.06.002
2014-01-01
Abstract:In the development of axial flow blood pump,the arterial partial flow field may produce an area with very low flow shear rate,so it is necessary to consider the non-Newtonian charac-teristics of blood fluid.In this paper,a model of axial flow blood pump was established,and flow and rotate-speed’s impacts on the inlet and outlet of the flow field in the blood pump were ana-lyzed through Computational Fluid Dynamics (CFD)simulation,as wel as the influence of the guide vane on the flow field.By the pump water experiment of the designed blood pump,its out-put flow and pressure were measured;the results show that the designed blood pump is consist-ent on the law with the simulation.
What problem does this paper attempt to address?